Applications, Measurement Channels, and Result Displays
R&S
®
FPL1000
89
User Manual 1178.3370.02 ─ 03
5.4.3 How to Set Up the Sequencer
In order to perform the configured measurements consecutively, a Sequencer function
is provided.
1. Configure a channel setup for each measurement configuration as required, includ-
ing the sweep mode.
2. In the toolbar, select the "Sequencer" icon.
The "Sequencer" menu is displayed.
3. Toggle the "Sequencer" softkey to "On" .
A continuous sequence is started immediately.
4. To change the Sequencer mode and start a new sequence immediately, select the
corresponding mode softkey, or press the RUN SINGLE or RUN CONT key.
The measurements configured in the currently active channel setups are per-
formed one after the other in the order of the tabs until the Sequencer is stopped.
The result displays in the individual channel setups are updated as the measure-
ments are performed.
To stop the Sequencer
► To stop the Sequencer temporarily, press the highlighted RUN SINGLE or RUN
CONT key (not for a channel-defined sequence). To continue the Sequencer,
press the key again.
To stop the Sequencer permanently, select the "Sequencer" icon in the toolbar and
toggle the "Sequencer" softkey to "Off" .
5.5 Channel Setup Overview
Each channel setup provides an overview of the most important currently defined set-
tings and access to the most important configuration dialog boxes for the particular
measurement.
Channel Setup Overview