Remote Commands
R&S
®
FPL1000
671
User Manual 1178.3370.02 ─ 03
Example:
INP:UPOR?
//Result: #B00100100
Pins 5 and 7 are active.
Usage:
Query only
8.8.5.2
Working with Power Sensors
The following commands describe how to work with power sensors.
These commands require the use of a Rohde
&
Schwarz power sensor. For a list of
supported sensors, see the data sheet.
.................................................................................. 671
Configuring Power Sensor Measurements
........................................................... 672
............................................................................. 679
Configuring Power Sensors
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe]
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
..........................................................671
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine
......................................................672
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe]
<State>
This command turns automatic assignment of a power sensor to the power sensor
index on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
SYST:COMM:RDEV:PMET:CONF:AUTO OFF
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
This command queries the number of power sensors currently connected to the
R&S
FPL1000.
Parameters:
<NumberSensors>
Number of connected power sensors.
Example:
SYST:COMM:RDEV:PMET:COUN?
Usage:
Query only
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters