![Rohde & Schwarz FPL1000 Series Скачать руководство пользователя страница 758](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826758.webp)
Remote Commands
R&S
®
FPL1000
758
User Manual 1178.3370.02 ─ 03
Example:
DEM:SQU:LEV 80
Sets the squelch level to 80% of the displayed signal.
[SENSe:]DEMod:SQUelch[:STATe]
<State>
This command turns selective demodulation at the marker position on and off.
For selective demodulation, the R&S
FPL1000 turns on a video trigger whose level
correponds to the squelch level. Therefore it turns other triggers or gates off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
DEM:SQU ON
Signals below the level threshold are not sent to the audio out-
put.
8.8.8.15
Programming Examples for Using Markers and Marker Functions
Various programming examples on how to use markers and the special marker func-
tions are provided here.
The use of spectrogram markers is demonstrated in
Example: Configuring a Spectrogram"
....................................................................................... 758
Example: Marker Search in Spectrograms
........................................................... 760
Basic Frequency Sweep Measurement for Marker Function Examples
Example: Using a Fixed Reference Marker
.......................................................... 761
Example: Obtaining a Marker Peak List
................................................................761
Example: Measuring Noise Density
...................................................................... 762
Example: Measuring Phase Noise
........................................................................762
Example: Measuring the Power in a Channel Using Band Power Markers
Example: Measuring Characteristic Bandwidths (Using the n dB Down Marker)
Examples: Demodulating Marker Values and Providing Audio Output
Example: Performing a Highly Accurate Frequency Measurement Using the Signal
Count Marker
Example: Basic Markers
This example demonstrates how to configure and define markers for a basic spectrum
measurement in a remote environment. It assumes that the basic frequency sweep
described in
Chapter 8.6.14, "Programming Example: Performing a Basic Frequency
on page 629 has been performed and thus does not begin by presetting the
instrument.
//--------------Configuring marker behavior ------------------
Setting Basic Spectrum RF Measurement Parameters