Chapter 8: Maintenance
5084773 A
215
18. Do one of the following to display the Configure Magnification dialog:
From the simplified menu, click Control > Devices > Configure Magnification.
From the standard menu, click Devices > Configure Magnification.
19. Enter the new objective information. See
Updating Magnification Settings on page 230
.
20. Do one of the following to display the Calibrate Distances dialog:
From the simplified menu, click Measure > Distances > Calibrate Distances .
From the standard menu, click Measure > Calibrate Distances.
21. In the Calibrate Distances dialog, verify that the pixel size calibration is set at or close to
the default value for this magnification. See
Calibrating Distances on page 231
22. Configure parfocality measurements using the IX Taskbar and enter X,Y Offset in the
Configure Magnification dialog.
If the IX Taskbar is unavailable, then use the bead plate to do parfocality measurements.
See
Configuring Parfocality After Changing Objectives on page 218
.
23. Use the IX Taskbar to measure pixel sizes, then enter the results in the Calibrate Distances
dialog. See
Calibrating Distances on page 231
24. Exit the MetaXpress software.
25. Start the Meta Imaging Series Administrator Software.
26. Enter the parfocality results (focus values) in the Objectives Settings dialog. See
the Focus Objective Values in the Meta Imaging Series Administrator Software on page
228
.
27. If you need to measure the plate bottom reference, then enable Maintenance Mode,
otherwise, exit the Meta Imaging Series Administrator Software and continue to the next
step. See
Finding the Plate Bottom Reference Objective and Reference Point on page
and
Editing Objective Settings in the Meta Imaging Series Administrator Software on
CAUTION! Unless there is an important reason to change the Plate Bottom
Reference Objective after the system is installed, you should leave the Plate
Bottom Reference Objective in place at all times. Contact Technical Support for
assistance.
28. Start the MetaXpress software.
29. Verify that the new objective is listed in the Magnification list. See
Dialog: Configure Tab, Objective and Camera Tab on page 69
.
30. If needed, create new shading correction files. See
Updating Legacy Shading Correction
31. Configure the Laser Auto Focus settings for the new objective with the plates and slides.
See
Confirming Laser Autofocus Settings for Plate Files on page 51
and
Setup Dialog: Configure Tab, Plate Tab on page 72
.
Содержание ImageXpress Confocal HT.ai
Страница 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Страница 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Страница 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Страница 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Страница 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Страница 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Страница 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...