ImageXpress Confocal HT.ai High-Content Imaging System User Guide
102
5084773 A
Option
Description
Notes
Off
No shading correction or background
subtraction is done.
Not recommended for montages.
Not recommended for assays where
intensity of individual objects will
be compared across the field of
view.
Auto
Correction for
FL
Combines FL Shading and FL
Subtraction. Two reference images
are created and used, one for the
shading and one for the subtraction.
Not recommended for Transmitted
Light images.
Not recommended for time lapse
assays of dim samples.
Not recommended for Z-stack
assays where 3D reconstruction is
used.
Not recommended for use with oil-
immersion objectives.
Low magnification objectives may
produce an artifact as it may catch
the edge of the plate in the
reference image.
Incorrect instrument configuration
could produce a stage error.
You should use the Auto Expose
tool when you optimize exposure
times or temporarily disable the
shading correction if you use snaps.
Turn back on to run Plate
Acquisition.
The subtraction portion of this
correction is not compatible with
water immersion objectives. A
message displays but acquisition
can continue and the shading
portion is still applied.
Table 5-10: W# Tab: Shading Correction Options
Содержание ImageXpress Confocal HT.ai
Страница 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Страница 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Страница 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Страница 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Страница 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Страница 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Страница 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...