Chapter 7: Expansion Solution Options
5084773 A
161
15. Reconnect the wires to the sensors on the green luer lock side.
Note: The color of the sensor wire is irrelevant when connecting to the sensor
socket. Either color of sensor wire can connect to either sensor socket.
Figure 7-34: Water Level Sensor Wire Sockets
Figure 7-35: Water Level Sensor Wires Inside the Instrument
16. Reconnect the CO
2
/air tubing to the luer locks on both sides of the water reservoir.
17. Replace the top access panel surrounding the top door and close the upper hinged side
door. See
18. Replace the lock screw in the top access panel. See
.
19. If your instrument includes the Transmitted Light option, move the Transmitted Light tower
back into position and replace the hinge release knob. See
20. Restart the system. See
Starting the System on page 27
21. In the MetaXpress software, do one of the following to close the instrument top door:
From the simplified menu, click Screening > Acquisition Setup > Load Plate.
From the standard menu, click Screening > Plate Acquisition Setup > Load Plate.
Содержание ImageXpress Confocal HT.ai
Страница 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Страница 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Страница 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Страница 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Страница 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Страница 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Страница 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...