Chapter 7: Expansion Solution Options
5084773 A
147
5. Remove the lock screw from the top access panel.
Figure 7-18: Top Access Panel Lock Screw
6. From the top of the instrument, lift the top access panel surrounding the top door and
remove it. Keep the instrument top door closed.
Figure 7-19: Top Access Panel Around the Top Door On and Off
7. Verify that there is a steady flow of bubbles in the water reservoir.
8. Replace the top access panel surrounding the top door.
9. Check the environmental control settings to verify that the CO
2
pressure is OK. See
Environmental Control Software on page 150
After Setting Up the CO
2
Tank
Load an unlidded plate with the live-cell sealing ring on top. If there are concerns about
contamination, a breathable seal can be used over the top of the plate.
Note: Before you do imaging experiments, wait for the system and plate to reach
equilibrium. Allow at least two hours for the system and 30 minutes for the plate.
Because focus settings and offsets change with temperature, you might need to
optimize them after the system and plate have reached equilibrium.
Содержание ImageXpress Confocal HT.ai
Страница 1: ...5084773 A March 2021 ImageXpress Confocal HT ai High Content Imaging System User Guide...
Страница 6: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 6 5084773 A...
Страница 16: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 16 5084773 A...
Страница 26: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 26 5084773 A...
Страница 62: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 62 5084773 A...
Страница 188: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 188 5084773 A...
Страница 248: ...ImageXpress Confocal HT ai High Content Imaging System User Guide 248 5084773 A...