Parameter
Description
Definition
Notes
Param Script
Visual Basic or Java script that
produces a measurement from
one or two input waveforms.
Available with XMAP,
XDEV, and ASDA options.
Standard in DDA-5005A.
Peak Mag
Peak mag away from a
baseline. Note: the measure
gate must include more of the
baseline than any other single
level.
Available
with
ENET
option.
Peaks
Number of peaks in a
histogram.
Available with DDM2, JTA2,
XMATH, XWAV, CAN02,
SDA, and XMAP options.
Standard in DDA-5005A,
SDA100G, and WaveExpert
scopes.
Peak to peak
Peak-to-peak: Difference
between highest and lowest
points in waveform. Unlike
ampl, does not assume the
waveform has two levels.
maximum - minimum
Gives a similar result when
applied to time domain
waveform or histogram of
data of the same waveform.
But with histograms, result
may include contributions
from more than one
acquisition.
Standard parameter.
Percentile
Horizontal data value that
divides a histogram so the
population to the left is xx% of
the total.
Available with DDM2, JTA2,
XMATH, XWAV, CAN02,
SDA, and XMAP options.
Standard in DDA-5005A,
SDA100G, WaveExpert, and
sampling scopes.
Per DCD
Amount of jitter due to duty
cycle distortion
Standard in SDA and
WaveExpert scopes.
Per Duty Cyc
Duty cycle measured on a
persistence
Standard in SDA and
WaveExpert scopes.
Period
Period of a cyclic signal
measured as time between
every other pair of 50%
crossings. Starting with first
transition after left cursor,
period is measured for each
transition pair, with values
averaged to give final result.
Where: Mr is the number of
leading edges found, Mf the
number of trailing edges
found,
the time when
rising edge i crosses the x%
level, and
the time when
falling edge i crosses the x%
level.
Standard parameter.
SDA-OM-E Rev H
169
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...