Group Delay Compensation
Group delay is defined as the rate of change of the total phase shift with respect to angular
frequency through a device or transmission medium. An unavoidable consequence of correcting
for group delay is that preshoot and overshoot will be added to the signal.
The SDA 18000/13000/11000/9000 provides a control in the channel "Vertical Adjust" dialog that
allows you to choose between two group delay compensations.
Pulse Response
- minimizes preshoot.
This selection is preferred in physics applications: no
"prediction" of the future event.
Eye Diagram
- less overshoot.
This selection improves the accuracy of serial eye diagrams.
Flatness
– like eye diagram.
This selection provides a flat frequency response.
SDA-OM-E Rev H
71
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...