Control Edge Triggering
Horizontal:
Turn the Delay knob in the HORIZONTAL control group to adjust the trigger's
horizontal position. Or, touch inside the
Delay
field in the timebase setup dialog and enter a
value, using the pop-up keypad.
The trigger location is shown by a marker under the grid:
Post-trigger delay is indicated by a left-pointing arrow below-left of the grid:
The time value is given in the title line of the
TimeBase
label below-right of the grid:
Vertical:
Turn the
L
EVEL
knob
in the TRIGGER control group to adjust the trigger's
vertical threshold.
Turn this knob to adjust the level of the trigger source or the highlighted trace. Level defines the
source voltage at which the trigger will generate an event a change in the input signal that
satisfies the trigger conditions.
Alternatively, in the "Trigger" dialog, you can touch inside the
Level
field and type in a value,
using the pop-up numeric keypad. To quickly set a level of zero volts, touch the
Zero Level
button directly below the
Coupling
field.
An arrow on the left side of the grid shows the threshold position. This arrow is only visible if the
trigger source is displayed.
Edge Trigger Setup
Channel Setup
1. In the menu bar, touch
Trigger
, and then select
Trigger Setup...
from the drop-down menu.
2. Touch
the
Edge
trigger button
under the
Trigger
tab.
80
SDA-OM-E Rev H
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...