SDA Operator’s Manual
5. Touch
inside
the
Compare Values
field and select
All
or
Any
from the pop-up menu
.
By selecting
All
, the test is true only if every waveform falls within the set limit. By selecting
Any
, the test is be true if just one waveform falls within the limit.
6. Touch
inside
the
Condition
field in the "ParamCompare" mini-dialog and select a math
operator from the pop-up menu:
7. Touch
inside
the
Limit
field and enter a value, using the pop-up numeric keypad. This value
takes the dimensions of the parameter that you are testing. For example, if you are testing a
time parameter, the unit is seconds. If you chose either
WithinDeltaPct
or
WithinDeltaAbs
from the
Condition
menu, you also have the choice of setting the
limit by means of the statistical buttons at the bottom of the "ParamCompare" dialog:
SDA-OM-E Rev H
207
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...