SDA Operator’s Manual
You can also use Sequence mode in remote operation to take full advantage of the instrument's
high data-transfer capability.
How the instrument captures segments
Sequence Mode Setup
Note:
Once a single acquisition has started, you can interrupt the acquisition at any time by pressing the Single front
panel button a second time. In this case, the segments already acquired will be retained in memory.
1. In the menu bar, touch
Timebase
, then touch
Horizontal Setup...
in the drop-down menu.
2. Touch
the
Smart Memory
tab, then touch the
Sequence
mode button
.
3. Under
Sequence Options
, touch inside the
Num Segments
data entry field and enter the
number of segments you want to display, using the pop-up keypad.
4. Touch
inside
the
Timeout
data entry field and enter a timeout value.
Note:
The timeout period accounts for instances when a
Num Segments
miscount occurs for some reason and the
scope waits indefinitely for an unforthcoming segment. During that time, no scope functions are accessible. By means of a
timeout value, however, the acquisition will be completed, the waveform displayed, and control of the scope returned to
the user after the timeout has elapsed.
5. Touch
the
Enable Timeout
checkbox.
6. In the menu bar, touch
Display
, then
Display Setup...
in the drop-down menu.
SDA-OM-E Rev H
63
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...