SDA Operator’s Manual
9. If you chose either
WithinDeltaPct
or
WithinDeltaAbs
from the
Condition
menu, touch inside the
Delta
field and enter a value.
Mask Testing
1. Touch
Analysis
in the menu bar, then
Pass/Fail Setup...
in the drop-down menu.
2. Touch
a
Qx
tab; a setup dialog for that position will open.
3. Touch
inside
the
Source1
field and select a source from the pop-up menu.
4. Touch
inside
the
Condition
field in the main dialog and select
Mask Test
.
5. From the "Test" mini-dialog, make a selection in the
Test is True when
group of buttons:
.
This selection means, for example, that if you select
All In
the test will be False if even a
single waveform falls outside the mask.
6. From
Show Markers
, choose whether or not to have mask violations displayed.
7. If you are loading a pre-existing mask, touch the
Load Mask
tab, then the
File
button. You
can then enter the file name or browse to its location.
8. If you want to make a mask from your waveform, touch the
Make Mask
tab.
9. Touch
inside
the
Ver Delta
and
Hor Delta
fields and enter boundary values, using the pop-up
numeric keypad.
10. Touch the
Browse
button to create a file name and location for the mask if you want to save
it.
11. Touch the
Gate
tab, then enter values in the
Start
and
Stop
fields to constrain the mask to a
portion of the waveform. Or, you can simply touch and drag the Gate posts, which initially are
placed at the extreme left and right ends of the grid.
SDA-OM-E Rev H
209
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...