Recalling Notebook Entries
After a notebook entry is made, you can recall it at any time. The recall includes waveforms and
scope settings.
1. Select the notebook entry from the list box.
2. Touch
Flashback
.
3. To exit Flashback, touch the
Undo Flashback
button
in the top-right
corner of the screen, or press the
Auto
trigger button.
Note
: The flashback feature currently recalls the DSO Setup, and all displayed waveforms. Some forms of ‘result data’ are
not recalled, including:
•
Persistence data. This will be saved in the hardcopy, and will be printed in the report, but will not be recalled during
Flashback.
•
Histogram data. Histograms internally have a 32-bit resolution, but when stored into a trace file and recalled during
flashback they are clipped to 16-bits.
•
Floating point waveforms. Certain math operations result in the creation of floating point waveforms with much higher
resolution than can be stored in a 16-bit waveform file. This extra resolution will not be preserved when traces are
recalled using flashback.
•
Cumulative Measurements. Any measurements that are on when the Lab Notebook entry is created are not saved
individually in the database, other than being embedded in the hardcopy image. This means that when flashback is
used, the measurements will be recomputed using the waveform data that was recalled. Normally this will not pose a
problem, but if cumulative measurements were on, which accumulated data from multiple acquired waveforms, they
will loose their history and show instead only the results from the stored waveforms.
298
SDA-OM-E Rev H
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...