168
SDA-OM-E Rev H
Parameter
Description
Definition
Notes
Minimum
Measures the lowest point in a
waveform. Unlike base, does not
assume waveform has two levels.
Lowest value in
waveform between
cursors
Gives similar result when
applied to time domain
waveform or histogram of data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition.
Standard parameter.
Nb phase
Provides a measurement of the
phase at a specific frequency of a
waveform (narrow band).
Available with DDM2, XMATH,
SDA, and XMAP options.
Standard in DDA-5005A and
SDA100G scopes.
Nb Power
Provides a measurement of the
power at a specific frequency of a
waveform (narrow band).
Available with DDM2, XMATH,
PMA2, SDA, and XMAP
options.
Standard in DDA-5005A and
SDA100G scopes.
N-cycle jitter
Peak-to-peak jitter between edges
spaced
n
UI apart.
Compares the
expected time to the
actual time of leading
edges n bits apart.
Available in SDA analyzers.
NLTS
Provides a measurement of the
nonlinear transition shift for a prml
signal.
Available with DDM2 option.
Standard in DDA-5005AA.
Num Points
Number of points in the waveform
between the cursors.
Standard parameter.
One Level
One level of an eye diagram
Available with SDA and SDM
options.
Standard in SDA, SDA100G,
and WaveExpert scopes.
Overshoot-
Overshoot negative: Amount of
overshoot following a falling edge,
as percentage of amplitude.
(base - min.)/ampl x
100
Waveform must contain at
least one falling edge. On
signals not having two major
levels (triangle or saw-tooth
waves, for example), may not
give predictable results.
Standard parameter.
Ov Overshoot
positive: Amount of
overshoot following a rising edge
specified as percentage of
amplitude.
(max. - top)/ampl x
100
Waveform must contain at
least one rising edge. On
signals not having two major
levels (triangle or saw-tooth
waves, for example), may not
give predictable results.
Standard parameter.
Overwrite
Ratio of residual-to-original power
of a low frequency waveform
overwritten by a higher frequency.
Available with DDM2 option.
Standard in DDA-5005A.
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...