SDA Operator’s Manual
Parameter
Description
Definition
Notes
Fall time
Fall time: Duration of falling edge
from 90-10%.
Threshold arguments specify two
vertical values on each edge used
to compute fall time. Formulas for
upper and lower values:
lower = lower thresh. x amp/100 +
base
upper = upper thresh. x amp/100 +
base
Time at upper
threshold minus
Time at lower
threshold
averaged over each
falling edge
On signals not having two
major levels (triangle or saw-
tooth waves, for example), top
and base can default to
maximum and minimum,
giving, however, less
predictable results.
Standard parameter.
Fall 80-20%
Fall 80-20%: Duration of pulse
waveform's falling transition from
80% to 20%, averaged for all falling
transitions between the cursors.
Average duration of
falling
80-20% transition
On signals not having two
major levels (triangle or saw-
tooth waves, for example), top
and base can default to
maximum and minimum,
giving, however, less
predictable results.
Standard parameter.
Fall@level
Fall at level: Duration of pulse
waveform's falling edges between
user-specified transition levels. See
also Rise@level.
Transition time for % or
absolute levels of all
falling edges.
Enhanced version sets
measurement
calculations to use one
of the following:
Base & Top (% or
absolute)
Peak-Peak (%)
0V-Min (%)
On signals not having two
major levels (triangle or saw-
tooth waves, for example), top
and base can default to
maximum and minimum,
giving, however, less
predictable results.
Standard parameter.
Enhanced parameter available
with EMC option.
First
Indicates value of horizontal axis at
left cursor.
Horizontal axis value
at left cursor
Indicates location of left
cursor. Cursors are
interchangeable: for example,
the left cursor may be moved
to the right of the right cursor
and first will give the location
of the cursor formerly on the
right, now on left.
Standard parameter.
SDA-OM-E Rev H
161
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...