TIMEBASE AND ACQUISITION SYSTEM
Timebase Setup and Control
Set up the timebase by using the front panel
Horizontal
controls, just as for analog scopes.
For additional timebase setups
1. Touch
Timebase
in the menu bar, then
Horizontal Setup...
in the drop-down menu. The
"Horizontal" dialog appears.
2. Touch
inside
the
Time/Division
data entry field and enter a value using the pop-up numeric
keypad, or use the up/down arrows to adjust the value.
3. Touch
inside
the
Delay
data entry field and type in a value, using the pop-up keypad. Touch
the
Set To Zero
button to set the delay to zero.
4. Touch
the
SMART Memory
button or tab and adjust the memory as needed.
Autosetup
When channels are turned on, Autosetup operates only on those turned-on channels. If no
channels are turned on, all channels are affected. When more than one channel is turned on, the
first channel in numerical order with a signal applied to it is automatically set up for edge
triggering.
You can perform an autosetup of all these functions together by simply pressing
on the
front panel, or by touching
Autosetup
in the Vertical, Timebase, or Trigger
drop-down menu.
SDA DBI Controls
The SDA 18000/13000/11000/9000 oscilloscopes achieve higher bandwidth and sample rate
using an innovative LeCroy technology called Digital Bandwidth Interleaving (DBI). This
technology allows resources to be borrowed from unused channels to multiply not only sample
rate but also bandwidth.
SDA18000 DBI Controls
When 18 GHz is selected, the active channels are either C1 or C4 or
both, providing 60 GS/s sampling on the 18 GHz channel, and 20 GS/s
sampling on the 6 GHz channel.
When 11 GHz is selected, the active channels are C1, C3 and C4. The
11 GHz channel provides 40 GS/s sampling, while the 6 GHz channels
provide 20 GS/s sampling.
When 6 GHz is selected, all 4 channels are active, each providing 20
GS/s.
72
SDA-OM-E Rev H
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...