SDA Operator’s Manual
SDA-OM-E Rev H
351
Control
Values
Description
Total Jitter
The total jitter curve which, shows the
cumulative probability of an edge being
greater (right-hand side) or less (left-hand
side) than its nominal location in the center.
The total jitter at a particular bit error rate is
measured from the width of this curve at the
desired bit error rate given by the vertical
location over which the width is measured.
EstTIE
Estimated tie histogram. This shows the
original histogram of TIE measurements that
has been extrapolated to a population of
10
16
.
LogTIE
The log of the TIE histogram
Bathtub
The bathtub curve. This is a an alternate
version of the total jitter curve scaled
horizontally to 1 UI. The right-hand side of
the curve is taken from the left-hand side of
the total jitter curve, and the left-hand side
comes from the right-hand side of the total
jitter curve.
Frequency
100 kHz to 4.5 GHz
The bit rate of the signal. This control is only
active when the bathtub output is active.
Transition density 0.1 to 1.0
The transition density in the data signal.
Usually assumed to be 0.5. This represents
the number of bit transitions compared to the
total number of bits in the data stream. This
value is used to normalize the extrapolated
histogram.
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...