Parameter
Description
Definition
Notes
Area
Integral of data: Computes area of
waveform between cursors relative
to zero level. Values greater than
zero contribute positively to the
area; values less than zero
negatively.
Sum from first to last
of data multiplied by
horizontal time
between points
Standard parameter..
Avg Power
Average power of the waveform
Available with SDA and SDM
options.
Standard in SDA100G scopes.
Base
Lower of two most probable states
(higher is top). Measures lower
level in two-level signals. Differs
from min in that noise, overshoot,
undershoot, and ringing do not
affect measurement.
Value of most
probable lower state
On signals not having two
major levels (triangle or saw-
tooth waves, for example),
returns same value as min.
Standard parameter.
Bit Rate
One over duration of one UI
measured on an eye
Available
with
SDA
option.
Standard in SDA, SDA100G,
and WaveExpert scopes.
CMACp
PCI Express V TX-CM-ACp and V
RX-CM-Acp
Available
with
PCIE
option.
Cycles
Determines number of cycles of a
periodic waveform lying between
cursors. First cycle begins at first
transition after the left cursor.
Transition may be positive- or
negative-going.
Number of cycles of
periodic waveform
Standard parameter.
cyclic
Mean
Cyclic mean: Computes the
average of waveform data.
Contrary to mean, computes
average over an integral number of
cycles, eliminating bias caused by
fractional intervals.
Average of data
values of an integral
number of periods
Choose this parameter by
selecting
Mean
from the
parameter table, then touching
the
Cyclic
checkbox.
Standard parameter.
Area
Integral of data: Computes area of
waveform between cursors relative
to zero level. Values greater than
zero contribute positively to the
area; values less than zero
negatively.
Sum from first to last
of data multiplied by
horizontal time
between points
Standard parameter..
cyclic
RMS
Cyclic root mean square:
Computes square root of sum of
squares of data values divided by
number of points. Contrary to rms,
calculation is performed over an
integral number of cycles,
eliminating bias caused by
fractional intervals.
Where: v
i
denotes measured
sample values, and N =
number of data points within
the periods found.
Choose this parameter by
selecting
RMS
from the
parameter table, then touching
the
Cyclic
checkbox.
Standard parameter.
156
SDA-OM-E Rev H
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...