3. There are two crossing-level controls in the SDA main dialog: one for the data signal under
test and another for the reference clock (if the
Recover clock
checkbox is unchecked). The
crossing time used by both the jitter and eye pattern measurements is determined as the
absolute time at which the signal level crosses the selected threshold. Touch inside the
Data
Level Is
or
Clock Level Is
field and select either
Absolute
or
Percent
from the pop-up
menu.
4. If you selected
Absolute
, touch inside the
Abs level
field and enter a voltage value.
Alternatively, you can touch the
Find Data Level
or
Find Clock Level
button. If you selected
Percent, touch inside the
Percent level
field and enter a percentage value, using the pop-up
keypad. The level is automatically set to the selected percentage on each acquisition.
Note:
The absolute level setting should be used for jitter measurements over several acquisitions because the relative
setting can potentially remove jitter by tracking slowly, thereby varying level shifts that could occur between acquisitions.
5. Touch
inside
the
Signal Type
field and select a standard signal type from the pop-up menu.
The signal type defines the compliance masks and bit rates for the selected standard. If you
selected
Custom
from the
Signal Type
menu, touch inside the
Bit Rate
field and enter a bit
rate, using the pop-up keypad. By touching the
Find Bit Rate
button, you can have the bit
rate determined automatically. This button is also available when a standard data type is
selected.
SDA-OM-E Rev H
313
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...