SDA Operator’s Manual
List of Parameters
The following table describes the instrument parameters. Availability of some parameters
depends on the options installed. See the comments in the "Notes" column of the table.
Parameter
Description
Definition
Notes
100BT Fall
Fall time between 2 levels (upper-
base, base-lower) of a 3-level
signal (100BT)
Available
with
ENET
option.
100BT Rise
Rise time between 2 levels (Lower-
base, base-upper) of a 3-level
signal (100BT)
Available
with
ENET
option.
100BT TIE
Difference between the measured
and ideal times at level between
base and upper orlower levels of
100BT signal.
Available
with
ENET
option.
100BT Tj
Total jitter from a TIE at level
between base and upper or lower
levels of 100BTsignal.
Available
with
ENET
option.
ACSN Auto-correlation
Signal-to-Noise
provides a signal-to-noise ratio for
periodic waveforms.
Available
with
DDM2
option.
Standard in DDA-5005A.
AltNCycle
Alternate N-Cycle Plot. Timing of
the transitions in the data
waveform is measured for each
transition and plotted as a function
of the number of unit intervals over
which the timing is measured. The
N-cycle plot displays the mean or
standard deviation of the edge
placement in the waveform relative
to each other (data to data) or to a
reference clock (clock to data).
Available
with
ASDA
option.
Amplitude
Measures the difference between
upper and lower levels in two-level
signals. Differs from pkpk in that
noise, overshoot, undershoot, and
ringing do not affect the
measurement.
top - base
On signals not having two
major levels (such as triangle
or saw-tooth waves), returns
same value as pkpk.
Standard parameter.
Ampl asym
Amplitude asymmetry between
taa+ and taa-
1 |(taa+ - taa-)|/(taa+
- taa-)
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
SDA-OM-E Rev H
155
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...