Interval Trigger that triggers when the interval falls within the selected range:
t1 = range's lower time limit; t2 = range's upper limit. The broken upward-pointing arrow indicates a potential
trigger, while the bold one indicates where the actual trigger occurs on the positive edge within the selected
range.
Interval Trigger Setup
1. In the menu bar, touch
Trigger
, then
Trigger Setup...
in the drop-down menu.
2. Touch
the
Interval
trigger button
.
3. Touch
inside
the
Trigger On
data entry field and select a source on which to trigger.
Touch inside the
Level
data entry field and enter a value using the pop-up numeric keypad.
4. Select positive or negative slope.
5. Touch
the
LessThan
button and enter a pulse-width value in the
Upper Limit
data entry
field.
Or
touch the
GreaterThan
button and enter a value in the
Lower Limit
data entry field.
Or
touch the
InRange
button.
88
SDA-OM-E Rev H
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...