348
SDA-OM-E Rev H
AltNcycle Control Summary
Control
Values
Description
Synch to pattern On/off
Synchronizes the measurement to a
repeating data pattern (if present) in the
signal when this control is checked. The
timing is measured between the first
transition in the pattern and each
subsequent transition. For an N-bit pattern,
each transition is measured in succession
and measurements are accumulated by
shifting the measurement window by N bits
and repeating the measurement.
Ver. Output
Mean, sdev, DDmean, DDsdev These are the vertical units of the
measurement. Mean gives the average
position of edges for each value of N relative
to a reference clock at the bit rate. Sdev
gives the standard deviation for each edge
for each value of N relative to the reference
clock. DDmean gives the mean value of
each edge relative to the first edge
normalized to a unit interval for each N.
DDsdev gives the standard deviation for
each edge relative to the first edge
normalized to a UI for each N.
Length (UI)
The maximum N value
This field displays the pattern length if
Sync
to Pattern
is selected. Otherwise, this is the
maximum N value displayed in the plot.
Tx dens.
0.0 to 1.0
Transition density of the data stream. This
control reads out the measured transition
density of the signal when the Find Pattern
button is pressed. This field is not used by
the N-cycle plot.
Seq. Type
PRBS or Pattern
Selects whether the repeating pattern is a
standard PRBS or a general pattern. This
control is only used when snyc to pattern is
selected.
PRBS
5 to 15
When PRBS is selected, sets the length.
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...