SDA Operator’s Manual
Parameter
Description
Definition
Notes
Jitter Filter
Jitter in the specified frequency
band. Generates a time sequence
of jitter measurements that are
filtered by the selected band-pass
filter.
Available
with
ASDA
option.
Last
Time from trigger to last (rightmost)
cursor.
Time from trigger to
last cursor
Indicates location of right
cursor. Cursors are
interchangeable: for example,
the right cursor may be moved
to the left of the left cursor and
first will give the location of the
cursor formerly on the left, now
on right.
Standard parameter.
Level@X
Gives the vertical value at the
specified
x
position. If the
x
position
is between two points, it gives the
interpolated value. When the
Nearest point
checkbox is
checked, it gives the vertical value
of the nearest data point.
Standard parameter.
Local base
Value of the baseline for a local
feature.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
Local bsep
Local baseline separation, between
rising and falling slopes.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
Local max
Maximum value of a local feature.
If 25% of the way
down toward min is
more than 10,000
points, the extremal
value is used as
the local max.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
Local min
Minimum value of a local feature.
If 25% of the way
up toward max is
more than 10,000
points, the extremal
value is used as
the local min.
Hysteresis argument used to
discriminate levels from noise
in data.
Available with DDM2 option.
Standard in DDA-5005A.
SDA-OM-E Rev H
165
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...