DSO Process
This instrument can be used to generate histograms of the parameter values of input waveforms.
The follwing items must be defined before doing so:
•
The parameter to be histogrammed
•
The trace on which the histogram is to be displayed
•
The maximum number of parameter measurement values to be used in creating the
histogram
•
The measurement range of the histogram
•
The number of bins to be used
Some of these items are pre-defined, but can be changed. Once defined, the oscilloscope is
ready to make the histogram. The sequence for acquiring histogram data is as follows:
1. Trigger
2. Waveform
acquisition
3. Parameter
calculations
4. Histogram
update
5. Trigger
re-arm
If you set the timebase for non-segmented mode, a single acquisition occurs prior to parameter
calculations. However, in Sequence mode an acquisition for each segment occurs prior to
parameter calculations. If the source of histogram data is a memory, saving new data to memory
effectively acts as a trigger and acquisition. This only takes place one time per second because
updating the screen can take lots of processing. This also minimizes trigger dead time. Under
remote control operation, the display can be turned off to maximize measurement speed.
Parameter Buffer
The oscilloscope maintains a circular parameter buffer of the last 20,000 measurements made,
including values falling outside the set histogram range. If the maximum number of events to be
used for the histogram is a number `N' less than 20,000, the histogram is continuously updated
with the last `N' events as new acquisitions occur. If the maximum number is greater than 20,000,
the histogram updates until the number of events is equal to `N.' Then, if the number of bins or
the histogram range is modified, the scope uses the parameter buffer values to redraw the
histogram with either the last `N' or 20,000 values acquired (whichever is the lesser value). This
allows the parameter buffer to redisplay histograms using an acquired set of values and settings.
This produces a distribution shape with the most useful information.
In cases where the optimal range is not apparent, the scope has a powerful range finding
function. If required, the values in the parameter buffer are examined to calculate an optimal
range and redisplay the histogram. The instrument also provides a running count of the
parameter value amounts falling within, below, or above the range. Values falling below or above
the range can then be recalculated by the range finder to include these parameter values (as long
as they are still within the buffer).
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...