30
SDA-OM-E Rev H
Memory
•
SDA 6020/4020:
20 Mpts/Ch. Option SDA-XL: 25 Mpts/Ch, 50 Mpts/Ch for 2 or 1
•
Ch.SDA6000A XXL/SDA4000A XXL:
50 Mpts/Ch, 100 Mpts/Ch for 2 or 1 Ch
•
SDA3010:
10 Mpts/Ch, 20 Mpts/Ch for 2 or 1 Ch. Option SDA-XL: 25 Mpts/Ch, 50 Mpts.Ch
for 2 or 1 Ch.
SDA 18000/13000/11000/9000
SDA 18000
SDA 13000/11000
SDA 9000
18 GHz/Ch
Mode
11 GHz/Ch
Mode
6 GHz/Ch
Mode
13/11 GHz/Ch
Mode
6 GHz/Ch
Mode
9 GHz/Ch
Mode
6 GHz/Ch
mode
Channels
1 1 4 2
4
2 4
Std Memory
60M 40M 20M 40M
20M
40M
20M
XL Memory Option
150M 100M 50M 100M 50M 100M 50M
Acquisition Modes
•
Random Interleaved Sampling (RIS):
200 GS/s for repetitive signals (SDA
18000/13000/11000/9000: RIS mode is available in 6 GHz mode only)
•
Single-Shot:
For transient and repetitive signals: 20 ps/div to 1000 s/div; (SDA
18000/13000/11000/9000 in 6 GHz mode only): 20 ps/div to 1 µs/div
•
Sequence:
2 to 40,000 segments (number of segments depends upon memory options)
•
Intersegment Time:
6 µs (typical)
Acquisition Processing
•
Averaging:
Summed averaging to 1 million sweeps; Continuous averaging to 1 million
sweeps
•
Enhanced Resolution (ERES):
From 8.5 to 11 bits vertical resolution
•
Envelope (Extrema):
Envelope, floor, roof for up to 1 million sweeps
Triggering System
•
Modes:
Normal, Auto, Single, and Stop
•
Sources:
Any input channel, External, EXT X10, EXT/10, or line; slope and level unique to
each source (except line trigger; disabled channels cannot be used as a trigger source in
SDA 18000, SDA13000, SDA 11000, or SDA 9000 when operating in 18 GHz, 13 GHz, 11
GHz, or 9 GHz modes.)
•
Coupling Modes:
DC
•
Pre-trigger delay:
0 to 100% of memory size (adjustable in 1% increments)
•
Post-trigger delay:
smaller of 0 to 10,000 divisions or 86400 seconds
•
Hold-off by time or events:
Up to 20 s or from 1 to 99,999,999 events
•
Internal trigger range:
±4.1 div from center
•
Max. trigger frequency:
Up to 5 GHz with Edge Trigger; 750 MHz with SMART Trigger
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...