182
SDA-OM-E Rev H
You can also enter combinations of the above units following the SI rules:
•
for the quotient of two units, the character
/
should be used
•
for the product of two units, the character
.
should be used
•
exponents can be represented by a digit appended to the unit without a space
For example,
•
acceleration can be entered as M/S2 for meters per second squared
•
volts seconds can be entered as V.S
In some cases, the units entered may be converted to simple units. For example entering V.A will
display W (watts)
Rescaling Setup
1. In the menu bar, touch
Math
, then
Math Setup...
in the drop-down menu.
2. Touch a math function trace tab
F1
through
Fx
The number of math traces available depends
on the software options loaded on your scope. See Specifications..
3. Touch
the
single
function button.
4. Touch
inside
the
Source1
data entry field and select a source: channel, math trace, memory
location.
5. Touch
inside
the
Operator1
data entry field and select
Rescale
from the
Functions
category.
6. In the dialog on the right, touch the
Rescale
tab.
7. Touch
inside
the
First multiply by
checkbox and enter a value for
a
, the multiplication factor.
8. Touch
inside
the
then add:
data entry field and enter a value for
b
, the additive constant.
9. Touch
inside
the
Override units
checkbox to disregard the source waveform's units, using
the pop-up keyboard.
Averaging Waveforms
Summed vs. Continuous Averaging
For Summed averaging, you specify the number of acquisitions to be averaged. The averaged
data is updated at regular intervals and presented on the screen.
On the other hand, Continuous averaging (the system default) helps to eliminate the effects of
noise by continuously acquiring new data and adding the new waveforms into the averaging
buffer. You determine the importance of new data vs. old data by assigning a weighting factor.
Continuous averaging allows you to make adjustments to a system under test and to see the
results immediately.
Note:
Continuous Averaging is accessible from the channel "Vertical Adjust" dialog under "Pre-Processing," and from the
math function menu.
Содержание SDA
Страница 1: ...SERIAL DATA ANALYZER OPERATOR S MANUAL December 2007 ...
Страница 148: ...Standard Horizontal Parameter Help Markers Standard Vertical Parameter Help Markers 148 SDA OM E Rev H ...
Страница 223: ...SDA Operator s Manual Example 6 SDA OM E Rev H 223 ...
Страница 225: ...SDA Operator s Manual SDA OM E Rev H 225 ...
Страница 232: ...In this figure the panel has been pasted onto the Excel sheet for comparison 232 SDA OM E Rev H ...
Страница 243: ...SDA Operator s Manual This example used the CORREL Array1 Array2 function of Excel as depicted below SDA OM E Rev H 243 ...
Страница 246: ...246 SDA OM E Rev H ...
Страница 247: ...SDA Operator s Manual Excel Example 5 Using a Surface Plot SDA OM E Rev H 247 ...
Страница 279: ...SDA Operator s Manual Convolving two signals SDA OM E Rev H 279 ...
Страница 310: ...The jitter wizard is accessed from the Analysis drop down menu 310 SDA OM E Rev H ...
Страница 327: ...SDA Operator s Manual SDA OM E Rev H 327 ...
Страница 328: ...328 SDA OM E Rev H ...
Страница 394: ...394 SDA OM E Rev H ...
Страница 395: ...SDA Operator s Manual SDA OM E Rev H 395 ...