Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
89
Symptoms
Possible Cause(s)
Corrective Action(s)
EMG response is
too high and
channel saturated?
Electrodes not picking up
the proper muscle
group?
Move either one or both measuring
electrodes (the ones attached to the
red & black leads) closer together.
EMG reading is too
high or channel
saturated?
Mains frequency pickup? Check the measurement electrodes and
quality of patient ground.
Pressure
Pressure not
responding?
Position of catheters?
Check position and adjust as
necessary.
Catheters are not
securely connected to
the PIM NXT?
Check connections and adjust as
necessary.
Zeroes not set properly? Reset zeroes in Preview workflow step.
Catheter kinked?
Replace catheter if necessary.
Pump
Pump not running? Pump pressure limit
reached?
Reset Pves to zero. Check for kinked
bladder catheter. Ensure Pump
pressure does not exceed 150 ml/min.
Pump settings?
Check Pump and Infusion lines. Check
if Pump door is closed. Check if Pump
tubing is inserted correctly.
Urocap
Uroflowmeter signal
shows vibration
and/or spike
patterns?
Plastic beaker is touching
the flow funnel?
Reposition the Uroflowmeter and re-
check.
Patient has touched the
Uroflowmeter with their
feet?
Ask patient to remain calm during
procedure.
Floor is unsteady?
Move to a more solid foundation.
Urocap NXT is recording
environmental pressure
changes?
Unplug the vent plug from the Urocap
NXT air vent located on the bottom of
the unit.
Activate the
Urocap Noise Reduction
Level
. Click
Settings
>
Workflow
>
Editor
to launch the
Workflow