Hardware
NXT Owner’s Manual, NXT-UM01, MAN2010
101
Figure 82 indicates the location of key controls on the Urocap NXT:
Figure 82: Urocap NXT Controls
Urocap NXT—Charging the Battery
To charge the Urocap NXT, place the device on the Urocap NXT Charging Hub, located on
the cart. A beeping sound signifies the Urocap NXT is connected. If the battery of the
Urocap NXT is completely drained, the device will not beep when connected to charge.
The table below indicates the status of the Urocap NXT and its battery.
NOTE:
For a visual reference of the Urocap NXT Charging Hub, refer to Figure 85.
Urocap NXT Status
Urocap NXT
Green: Ready to use
Yellow: Not ready to use, commence charging
Lack of Illumination: Device inactive or in sleep
mode; not ready to use.
Device will go into sleep mode one hour after
system shutdown. Return the Roam NXT to the
docking station for charging or activation.
Table 12: Urocap NXT LED Signals
LABORIE recommends for the user to have at least 1-2 hours of battery charge on the
Urocap NXT prior to the start of a test or to complete the test with the Urocap NXT plugged
in using the NXT Magnetic Pogo Pin Cable. Always return the device to the Urocap NXT
Charging Hub at the completion of working hours to ensure full charge for next study
period.
Urocap NXT Charging Guidance:
The device battery shall charge from “low” to “charged” in less than 5 hours when in charge
only mode or within 8 hours if charging while acquiring data. “Charged” is battery capacity
sufficient for more than or equal to one hour of recording and “low” is battery capacity not
sufficient for one hour of recording.
Reset Button
Air Vents
Charging Port
Vent Plug