Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
75
Wipe the area with a dry, lint free cloth. Ensure the area is dry to the touch.
Reattach, if removed, and close the transducer doors.
The transducer doors should only be kept open when reprocessing the area normally
covered by the transducer doors. Once reprocessing of this area is complete, close the
transducer doors to ensure that the transducer channels are not exposed to any liquids
during any additional processing.
5.1.3.4 Cleaning the Infusion Pump NXT
To clean the outer surface area of the Infusion Pump NXT, follow the instructions provided
in the
Performing Manual Cleaning
section. To clean the rollers and drip tray, complete the
following steps:
1.
Open the pump door and wipe each roller with an approved cleaning reagent. Refer
to the
Cleaning and Disinfection Reagents
section on page
71
.
2.
Remove the drip tray from the pump (Figure 61) and clean it separately with an
approved cleaning reagent.
Figure 61: Pump Rollers and Drip Tray
5.1.3.5 Performing Manual Cleaning
This section covers the cleaning method for the Cart, Shelves, Docking Station, Remote
Control NXT, and peripherals.
For the Roam NXT, PIM NXT, Urocap NXT, and NXT Infusion Pump, ensure to refer to the
device specific cleaning sections starting on page
73
for guidance on when and where to
apply the cleaning procedure described below.
Ensure to review
General Cleaning and Disinfection
,
Cleaning and Disinfection
Preparation
,
Cleaning
, and
Completing Cleaning and Disinfection
sections fully
before beginning cleaning.
Turn off and disconnect all power to the NXT System or NXT module.
Remove any visible soiling by hand immediately after the completion of a study
using a dry, soft, lint free cloth.
Gently wipe down the surface of the Cart, Shelves, Computer, NXT module or
peripherals using a fresh cleaning wipe.