How to Perform a Study
NXT Owner’s Manual NXT-UM01, MAN2010
33
Caution:
Infusion can be stopped at ALL TIMES by pressing the grey pump stop button
on the NXT Docking Station, by opening the pump door, or by disconnecting the filling tube
from the catheter or the water bag.
3.3.2.4 Voiding Phase
The
Voiding
phase will begin (Figure 19). Click the
Permission To Void
button
and ask the patient to void through the commode chair into the Urocap NXT
setup. See Preparing the Urocap NXT for a reference image of the Urocap NXT
Setup. If the patient fails to void re-click the
Permission to Void
button to
create a new marker; this action will shift the previous marker to ‘Attempt to
Void’.
Figure 19: Pressure Flow Voiding Phase
The voiding phase will stop 60 seconds after voiding ends.
A Post Voided Residual
window (Figure 16) will pop-up. Select the preferred
method of capturing Post Voided Residual (PVR), meaning the amount of fluid
remaining in the bladder at the completion of the voiding phase. Click
OK
to
move to the Report.
3.3.2.5 Urology Urodynamics Report
A preview of the Pressure Flow Report will be generated using the default (UDS)
configuration settings. The user can apply the default configuration for the report
or customize the report. For information on customizing the report, refer to the
Reports
section on page
60
.
Click the
Finish
button to close the report. The report will then appear under the
Studies
tab in the
Patient Details
section.
The operator may begin a secondary test with the same patient, under the
guidance of the overseeing physician, or continue to finish the study.