Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
84
5.3.2.2 Verifying Pressure Measurement Capability
To begin verifying calibration for the PIM NXT transducers, gather one Cylinder at least 25
cm tall and one T-DOC Air Charged Catheter. To complete calibration verification:
a)
Fill a cylinder with 20cm of water.
b)
Place a T-DOC Air Charged Catheter inside the cylinder so that the tip of the catheter
is touching the bottom of the cylinder.
c)
Insert the catheter connectors into the relevant PIM NXT transducer channels.
d)
Click the
Settings
icon in the title bar, then click the
System Settings > Device
Manager
option. The
Device Manager
window will pop-up.
e)
Click the
Roam NXT
option, then click the
Run Diagnostics
button. The device
diagnostics window will pop-up (Figure 67).
Figure 67: Roam NXT – Device Diagnostics
f)
The Pressure P1 should be approximately 18cm of water +/-3%.
Infusion Pump NXT and Volume Transducer –
Verifying Calibration
To set up verification of calibration for the Infusion Pump NXT and Infusion Volume
Transducer:
1.
Login to the Synergy software.
2.
Ensure the Infusion Pump NXT is connected to the Synergy software.