Quick Start
NXT Owner’s Manual NXT-UM01, MAN2010
18
Quick Start
To prepare for a typical study, follow the steps provided below.
Setting Up the Equipment
Gathering Supplies and Equipment
Preparing the Urocap NXT
Preparing the Roam NXT
Setting Up Consumables and Preparing the Patient
2.1 Setting Up the Equipment
Unpack the NXT Cart, Roam NXT, Patient Interface Module NXT (PIM NXT), and
Urocap NXT from their shipping boxes. For information on the NXT Cart, Roam
NXT, PIM NXT, Urocap NXT, Infusion Pump NXT, or Remote NXT, refer to the
Hardware
section
on page
93
.
IMPORTANT:
Lock all wheels of the NXT Cart for equipment setup and anytime the
NXT System is not actively being transported. Ensure all wheels are
locked during testing procedures to prevent injury to the patient.
To lock the wheels, press down on the lever with the bottom of your
foot as shown in Figure 2. To unlock, push the lever up from
underneath with the top of your foot.
Connect the PIM NXT to the Roam NXT. For instructions on how to connect these
devices, refer to the
Replacing the PIM NXT
section on page
98
.
If utilizing Video function for Urodynamic study prepare the connecting device.
Synergy supports HDMI and DVI connections provided by the AIO PC. BNC
connection is supported with the use of the supplied (where applicable) DVI to
BNC adaptor.
CAUTION:
Consult LABORIE Service before connecting any video device to the NXT
System. Anyone connecting supplementary equipment to ME equipment configures a
medical system and is therefore responsible for ensuring that the system complies with the
requirements for medical electrical systems, encompassed in the IEC/EN 60601 series.
Insert the Roam NXT into the NXT Docking Station and the Urocap NXT into its
Charging Hub, located on the NXT Cart, as shown in the figure below.
Figure 2: Lockable
Wheels