Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
85
5.3.3.1 Checking the Total Weight without a Saline Bag
To begin calibration verification, measure the total weight of the Infusion Pump NXT and
Infusion Volume Transducer without a saline bag:
a)
Setup the Infusion Volume Transducer without a saline bag.
b)
Click the
Settings
icon in the title bar, then click
System Settings
>
Device
Manager
option. The
Device Manager
window will pop-up.
c)
Click the
Infusion Pump NXT
option, then click the
Run Diagnostics
button. The
device diagnostics window will pop-up (Figure 68).
d)
Press the
Zero
button.
e)
The Total Weight should measure as zero immediately after zeroing. To complete
calibration verification, check the total weight with a saline bag.
Figure 68: Infusion Pump NXT Diagnostics Window
5.3.3.2 Checking the Total Weight with a Saline Bag
To complete calibration verification, check the total weight of the Infusion Pump NXT and
Infusion Volume Transducer with a saline bag:
a)
Press the
Zero
button.
b)
Setup the Infusion Volume Transducer with a saline bag of weight 500ml.
c)
The Total Weight in the diagnostic window must be 500.0g +/- 5%.