Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
76
Use a fresh wipe, or turn the wipe over to the clean side, to thoroughly wet the
surface of the device.
Wipe around cables, edges and corners. Ensure that all surface areas and
connectors have been in contact with the wipe.
Repeat as necessary until surfaces are visibly clean. If visible soil persists, use a
soft bristle brush to gently brush away or loosen the remaining soil, then wipe the
device again using a fresh wipe.
Wipe the Cart, Shelves, Computer, modules, and peripherals with a dry, lint free
cloth. Ensure the device is dry to the touch before using the device in a procedure.
Visually inspect the units in standard office lighting to ensure they are clean. If the
inspection determines the devices to not be visually clean, the user should repeat
the cleaning cycle.
Perform visual inspection of all units for corrosion, discoloration, pitting, or
cracking. Contact LABORIE for service and/or disposal instructions if the unit is
observed with these defects.
NOTE
: There are no rinsing steps required for the cleaning or disinfection processes.
Disinfection
5.1.4.1 Performing Manual Disinfection
To disinfect the device, follow the cleaning instructions allowing the surface to
remain wet for the required contact time as listed on the product label of
the cleaning solution. Ensure to refer to module specific cleaning instructions.
Ensure the device is dry to the touch before using it to perform another study.
Once disinfection is complete, confirm that each module is working correctly by
verifying calibration. Refer to the
Verifying Calibration
section on page
81
for
calibration instructions per module.
IMPORTANT
: Refer to Urocap NXT specific disinfection instructions in the
Disinfecting
the Urocap NXT
section on page
76
. Refer to Roam NXT and PIM NXT specific disinfection
instructions in the
Disinfecting the Roam NXT and PIM NXT
section on page
77
.
5.1.4.2 Disinfecting the Urocap NXT
To disinfect the Urocap NXT, follow the instructions provided in the
Performing Manual
Cleaning
section on page
75
allowing the surface to remain wet for the required contact
time listed on the product label of the cleaning solution.
Disinfect the area around the battery cover ventilation holes on the Urocap NXT following
the instructions below:
Remove the vent plug if connected.
Gently wipe the affected area with a cleaning wipe to remove soiling and debris.
Take care to prevent any excess cleaning reagent from entering the vent hole.
Allow the surfaces to remain wet for the required contact time listed on the
product label of the cleaning solution.
Repeat this process, if necessary, using either a fresh wipe or the clean side of the
existing wipe until the affected area is visibly clean.
Re-insert the vent plug.