Table 11: Roam NXT LED Signals
Roam NXT Charging Guidance:
The device battery shall charge from “very-low” to “charged” in less than 5 hours when in
charge only mode or within 8 hours if charging while acquiring data. “Charged” is battery
capacity sufficient for more than or equal to two hours of recording and “very-low” is
battery capacity sufficient for at least one-half hour of recording.
6.7 Urocap NXT
The Urocap NXT with wireless technology (Figure 81) is a portable Uroflowmeter that
provides a non-invasive and accurate measurement of urinary voiding functions.
•
The Urocap NXT is intended to be used to quantify the flow characteristics of the
lower urinary tract. Using a built-in transducer, the system performs standard
Uroflowmetry studies that measure both flow rate and voided volume.
Figure 81: Urocap NXT
•
The Bluetooth technology built into the Urocap NXT provides the patient with
more privacy during voiding. Bluetooth connection allows the Urocap NXT, the
beaker, the commode chair, and the funnel to be set up in one room while the
cart, computer, printer, and study operator can be in another room collecting and
printing data.
•
The Urocap NXT features air vents along its base. The center air vent should be
plugged during cleaning to avoid fluid ingress. During a study the plug should be
removed to allow external and internal pressures to equalize.
Solid Amber: Low (Battery capacity
sufficient for at least 1 hour of
recording)
Amber: Not ready to use.
Return the Roam NXT to docking
station for charging.
Flashing Amber: Very Low (Battery
capacity sufficient for at least 1/2
an hour of recording)
Lack of Illumination: Empty/Sleep
mode (No declared battery
capacity)
Lack of Illumination: Device inactive
or in sleep mode; not ready to use.
Device will go into sleep mode 1