Quick Start
NXT Owner’s Manual NXT-UM01, MAN2010
25
IMPORTANT:
TDOC Air-Charged Catheters are intended for single-use. Synergy will
not recognize a catheter that has previously been registered in the software after the
allotted two hours to complete Urodynamic testing.
Once preferred catheter placement is confirmed, connect the Infusion Pump
Tubing to the filling lumen on the vesical catheter or filling catheter. Ensure all air
bubbles have been removed from the Infusion Pump Tubing and that the Infusion
Pump has been properly primed before connecting the line to the filling lumen on
the vesical catheter or filling catheter.
Refer to the
T-DOC Air-Charged Catheters
section on page
104
for information on T-DOC
catheters.
Placing the EMG Patch Electrodes
Enter consumables information into the
Consumable Traceability
section in the
Preview
phase.
Place two electrodes peri-anally at the 10 o’clock and 2 o’clock position (or at the
9 o’clock and 3 o’clock position), as close to the anal verge as possible.Place the
third electrode on a bony prominence (such as the knee) or on a fatty portion of
the inner thigh.
Figure 9: EMG Patch Electrode Placements
Gather all tubing/wires and secure them to the patient’s thigh with tape.
Connect the EMG cable to the PIM NXT as shown in Figure 10 below.
Figure 10: EMG Cable Connected to PIM NXT
Performing Roam NXT Patient Setup
For ease of movement and accessibility during Urodynamic procedures, the Roam NXT may
be mounted on an IV Pole or on the procedure chair. When mounting the Roam NXT,
always ensure the clamping mount is securely fastened.
Figure 11: Roam Mounted on IV Pole