Introduction
NXT Owner’s Manual NXT-UM01, MAN2010
16
LABORIE is not responsible for loss of patient files or test data. LABORIE recommends
that the user backup patient data on a regular basis.
To prevent any risk of damage to the cart:
o
Ensure the monitor is free of any obstacles.
o
Do not hang anything on, or suspend extra weight from, the monitors.
o
Do not exceed the maximum weight indications: 10kg for the printer shelf,
10kg for the worksurface, and 1 kg for the mouse tray.
LABORIE equipment and accessories are licensed by Governments, approved by Safety
Agencies, and warranted to work only with each other.
The Roam NXT and the PIM NXT are not intended for long-term direct patient contact. If
the devices must be attached to a patient, place gauze or another material between the
devices and the patient’s skin.
Caution:
United States Federal Law restricts this device to sale or use by or on the
order of a licensed physician.
Warnings:
Only LABORIE trained technicians may service the unit.
Do not use if the device packaging has been opened, damaged, or if it presents any fault
due to improper transport, storage, or handling that could in any way hamper its use.
The NXT System is intended for install by trained personnel ONLY.
Ensure all wheels are locked during testing procedures to prevent injury to the patient. Do
not lean on the equipment or hang excessively weighted objects on the platform; only
hang LABORIE standard accessories on the platform.
Exhibit caution when adjusting the PC and monitor from its lowest or highest position;
ensure all extremities are clear of moving parts. If the PC or monitor are removed from
the mount, the unit may rise abruptly as the gas spring triggers used are calibrated to
handle the weight of the original configuration.
Do not immerse the NXT System or any of its modules in water or any other liquids.
Do not place your fingers or any body parts inside the pump when the pump rollers are
moving.
Do not use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous
oxide.
Do not use electromyography (EMG) simultaneously with high frequency surgical
equipment.
Do not operate the NXT System near shortwave or microwave therapy equipment as
proximity may produce instability in the APPLIED PARTS.
Do not allow applied parts linked to a PATIENT CONNECTION, but not connected to the
PATIENT, to contact other conductive parts including those connected to protective earth.
Preserve the patient connection electrical isolation.
Do not reuse single-use accessories. After use, dispose of single-use accessories in
accordance with local regulations.