Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
86
5.4 Configuration
NXT PRO Electrical Isolation Diagram
Refer to the diagram shown below in Figure 69 for a visual representation of the NXT Pro
electrical isolation from mains supply. For further information regarding electrical isolation,
contact LABORIE Service.
HUB
Spine
Spine
Work Station
BASE
AC/DC
18V~24V
PIM1
PIM2
18V~24V
Roam Base 1
Roam Base 2
Pump
J
18V
Isolation Barrier2
Isolation Barrier1
100-240VAC
50/60Hz
Direct electrical con
Indirect electrical c
J
Docking Station
UroCap 6
Extension
cable
Figure 69: NXT Pro Isolation Diagram
Urocap NXT and Wall Charger Isolation Diagram
Refer to the diagram below in Figure 70 for a visual representation of the method to isolate
the Urocap NXT from supply mains when configured with a wall charger. For further
information regarding electrical isolation, contact LABORIE Service.
UroCap 6
Direct electrical contact
Indirect electrical contact
100-240VAC
50/60Hz
Isolation Barrier1
Isolation Barrier2
Urocap Power Supply
P/N:POW1033
Cable NXT POGO PIN to
BINDER 3 meter
P/N:CAB1081
Figure 70: Urocap NXT and Wall Charger Isolation Diagram