Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
72
5.1.1.2 Cleaning and Disinfection Frequency
•
Cleaning –
After each use, all devices should be visually inspected and cleaned
to remove any debris or soiling.
•
Disinfection –
Devices should be disinfected in accordance with the frequency
defined below.
o
Roam NXT, PIM NXT, Urocap NXT:
Perform disinfection after the device has
contact with the patient’s skin, after contamination has been observed (e.g.
presence of urine on the device), or, at minimum, daily.
o
Docking Station, Pump, PC:
Perform disinfection of the device, at minimum,
once a week.
5.1.1.3 Methods and Tools
•
All cleaning and disinfection should be performed using moistened cloths and a
soft bristle brush.
•
No additional tools are required for cleaning or disinfection.
•
Inspect the Cart, Shelves, Computer, NXT modules, and peripherals (such as
keyboards and computer mouse) for physical damage or wear and tear prior to
cleaning.
•
Ensure the charging contacts for all modules are wiped free of debris and
contaminants.
•
Optional reusable accessories used in conjunction with the NXT System should be
cleaned and/or disinfected according to the instructions contained in their device
specific Instructions for Use. Refer to the
Introduction
section on page
12
an
overview of standard equipment and accessories.
CAUTION:
No abrasive tools should be used. Do not spray the cleaning reagents on
the NXT System components, modules, accessories, or peripherals. Do not autoclave the
NXT System components, modules, accessories, or peripherals or clean them with sharp
tools or abrasives. Do not immerse or rinse the NXT System components, modules,
accessories, or peripherals. Do not immerse electrical connectors in water or other liquids.
Cleaning and Disinfection Preparation
To begin cleaning or disinfection preparation, complete the following:
Put on gloves and safety glasses.
Obtain the module for cleaning and select the cleaning wipes.
Put on any other specific personal protective equipment recommended by the
wipe’s manufacturer.
Ensure a stable surface is available to place any modules or parts that may need
to be separated from the NXT System during the cleaning and disinfection process
in the designated reprocessing area.
5.1.2.1 Containment and Transportation
Transport the soiled device individually in a closed container to a designated cleaning and
disinfection area.
•
Do not transport the device in a container with other devices.
•
Do not soak the device prior to cleaning.
Transport the cart by first adjusting the system into the transportation mode described in
the
Safely Moving the Cart
section on page
79
. To limit cross contamination, push the cart
slowly to the designated cleaning area and avoid contact with other objects.