Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
77
5.1.4.3 Disinfecting the Roam NXT and PIM NXT
To disinfect the Roam NXT and PIM NXT, follow the instructions provided in the
Performing
Manual Cleaning
section on page
75
allowing the surface to remain wet for the required
contact time listed on the product label of the cleaning solution.
Disinfect the area of the PIM NXT normally covered by the transducer doors following the
instructions below:
Remove the transducer doors and place safely aside.
Gently wipe the affected area with a cleaning wipe taking care to avoid the
transducer channels. Take care to prevent any excess cleaning reagent from
entering the transducer channels. Wipe each transducer door to remove soiling
and debris. Allow the surfaces to remain wet for the required contact time listed
on the product label of the cleaning solution.
Repeat this process, if necessary, using either a fresh wipe or the clean side of the
existing wipe until the affected area is visibly clean.
Reattach and close the transducer doors.
Completing Cleaning and Disinfection
Follow the instructions provided below to complete the cleaning and disinfection process.
5.1.5.1 Reconnect and Reassemble
•
If previously disassembled, reassemble the Roam NXT and PIM NXT by inserting
the PIM NXT into the Roam NXT collar until it is flush. Screw together re-using
the screw supplied with the device. Ensure not to overtighten the screw.
•
Reattach PIM NXT and Roam NXT belt clips. Refer to the
Removing and
Attaching the Clip
section on page
99
.
•
Ensure to return the drip tray to the pump and close the pump door.
To use the system after cleaning, re-connect power to the system, dock devices, and follow
the instructions provided for normal use.
5.1.5.2 Packaging and Storage
Devices may be packaged flat in a clean, anti-static box or wrapped in a lint free cloth post-
cleaning or disinfection. Do not stack devices on top of each other and ensure that devices
are not pressed against each other or against the container. If the box has multiple
units, use a divider to ensure no contact with other devices.
Store all NXT modules in their respective docs or according to packaging instructions. Refer
to the
Appendix
on page
106
for Storage Conditions.
For long term storage, LABORIE recommends that the Roam NXT and the Urocap NXT be
fully charged prior to storing. Remove the module from the applicable docking station and
press the “Reset” button on the module to cause the device to go into long-term sleep
mode. To wake up the modules after storage, place them in the applicable docking station.