Maintenance and Service
NXT Owner’s Manual, NXT-UM01, MAN2010
82
d)
Press the
Zero Weight
button.
e)
The
Absolute Weight
and
Relative Weight
for Urocap NXT should measure at
0.0g +/- 2g immediately following zeroing. To verify calibration, check the absolute
and relative weight with water.
5.3.1.2 Checking the Absolute and Relative Weight with Water
To complete calibration verification for the Urocap NXT, check the absolute and relative
weight with water:
a)
Take a beaker and measure its weight (W
b
) on the Urocap NXT. Press the
Zero
Weight
button.
b)
Add 500 ml of water to the beaker.
c)
Verify that the
Absolute Weight
and
Relative Weight
for the Urocap NXT in the
diagnostic window is 500.00 g +/-3% (W
b
+ Water (500ml)) as exemplified in Figure
65.
Figure 65: Urocap NXT – Device Diagnostics with Water
Roam NXT – Verifying Calibration
To setup verification of calibration for the Roam NXT:
1.
Login to the Synergy software.
2.
Ensure the PIM NXT is connected to the Roam NXT.
3.
Connect the Urocap NXT, Roam NXT, and the Infusion Pump NXT to the Synergy
software and setup the hardware channels.