Chapter 10 Care & Maintenance
10-17
NOTE:
It is not necessary to test each lead individually or power condition combinations as required in
previous tests.
CAUTION
Line voltage is applied to the ECG leads during this test. To avoid possible electric shock
hazard, the system being tested must not be touched by patients, users or anyone while the ISO
TEST switch is depressed.