LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 144
Section 8-8 - PC Block
8-8-3
SMPS Assy, Hard Disk, CD R/W (FRU No. 601, 602, 603)
Purpose: This is a description on how to remove and replace the Parts in the BEP Assy.
8-8-3-1
Tools
•
Common pillips screwdrivers
8-8-3-2
Needed Manpower
•
1 persons, 15 m travel
8-8-3-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Left Cover should be removed before proceed. Refer section 8-5-1 on page 41, section 8-5-13 on
8-8-3-4
Removal Procedure
1.) Open front base cover and front cover. Refer section 8-5-6 on page 51, section 8-5-7 on page 53.
2.) Remove PC top fix bracket and scan EMC bracket. Refer Figure 8-158.
3.) Open the BEP metal cover
Figure 8-158 ATX SMPS Replacement
S c a n E M C B ra c k e t
5
1
2
3
4
6
7
8
P C T o p F ix B ra c k e t
S c a n E M C B ra c k e t
5
1
2
3
4
6
7
8
P C T o p F ix B ra c k e t