8 - 104
Section 8-8 - PC Block
8-8-1-5 BEP Cable Connection with Console( In case of change BEP 1 into BEP2 or BEP3)
5.) Connect LI, LO, MI, ETHERNET Connector to Sound/LAN Port on BEP.
6.) Connect CON17 Connector to BEP.
7.) Check the CON 17 cable is arranged without short before connecting.
Figure 8-95 Connecting Cables
Figure 8-96 Connecting CON17 cables
Ethernet Cable
MI(Mic:Red) Cable
LO(Line Out : Green) Cable
LI(Line In:Blue) Cable
Ethernet Cable
MI(Mic:Red) Cable
LO(Line Out : Green) Cable
LI(Line In:Blue) Cable
C O N 1 7 c a b le
C O N 1 7 c a b le