Chapter 4 Functional Checks
4 - 35
4-11-10
Menu Key Assy
1.) Boot up the system and run the scfpapitest program.
2.) Press and Rotary sub keyboard, and Paddle(up/down, right/left) button.
3.) Check the log dialog box in scfpapitest program.
NOTICE
Follow up the procedure from step 1 to step 7 of Section 4-11-1 to run the scfpapitest program.
Button or Rotary Name
The Name displayed in Log
Patient
New Patient Button
Report
Report Button
End Exam
End Exam Button
Mode Parameters
Application Button
Utility
Utility Button
Spare button
L5Pro Spare Button
Up
Paddle Up Button
Down
Paddle Down Button
Left
Paddle Left Button
Right
Paddle Right Button
Table 4-27 Button and Rotary Table