GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207, R
EVISION
7
LOGIQ™5 PRO S
ERVICE
M
ANUAL
2-2
Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1
Console Environmental Requirements
2-2-1-1
Cooling
The cooling requirement for the LOGIQ™ 5 is 3500 BTU/hr. This figure does not include cooling needed
for lights, people, or other equipment in the room. Each person in the room places an additional 300
BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and diameters can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interface.
2-2-2
Electrical Requirements
NOTE:
GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Table 2-2
Environmental Requirements for LOGIQ™ 5 Scanners
Operational
Storage
Transport
Temperature
10 - 35
o
C
50 - 104
o
F
-10 - 60
o
C
14 - 140
o
F
-40 - 60
o
C
-40 - 140
o
F
Humidity
30 - 85%
non-condensing
30 - 90%
non-condensing
30 - 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Table 2-3
Environmental Requirements for an Ultrasound Room
Item
Values
Power Source
Refer to Table 2-4 on page 2-3.
Current Rating
15A (120V-100V); 8A (220-240V) CIRCUIT BREAKER
Radiation Shielding
NONE REQUIRED for ULTRASOUND ENERGY
Temperature
20-26 DEG. C (68-79 DEG F) for PATIENT COMFORT
Humidity
50% to 70% for PATIENT COMFORT
Heat Dissipation
2000 BTU/Hr.
Floor Landing
Approximately 680 - 800 kg/m
2
without Accessories
Floor Condition
Gradient: WITHIN 5 degrees
Weight
180 kg (397lbs) without Accessories
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...