LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 160
Section 8-9 - Power Block
8-9-3-5
Mounting procedure
Case1
Replacing JPC with JPC
Replacing JPC2 with JPC2
Install the new parts in the reverse order
Case2
Replace JPC with JPC2
Figure 8-178 JPC2 ASSY installation
CAUTION
J4 and J12 are not used with the JPC2, since JPC2 Assay has Relay instead of SSR and it does
not have to connect J4 ( SSR control connector ), J12 ( SSR power ) to JPC2.
J4 connector
( SSR Control signal )
J12 connector
( SSR power )
JPC2 assy