LOGIQ™5 PRO S
ERVICE
M
ANUAL
5-38
Section 5-9 - Air Flow
Section 5-9
Air Flow
5-9-1
Air Flow Distribution
The four air flow passes allow the scanner to be cooled down as shown in the figure above.
•
Pass A (Bottom right> BEP Assy> Rear upper left) for BEP Assy cooling.
•
Pass B (Front lower left > Filter > LV unit > Rear lower left) for HV unit cooling.
•
Pass C (Bottom left> RLY Box Assy > Rear upper right) for RLY Box Assy cooling.
•
Pass D (Bottom left> Filter >Nest Assy> Rear upper right) for Nest Assy cooling.
5-9-2
Filters
The scanner contains the four filters located at:
-
Front bottom for air flow of the BEP Assy and Nest Assy.
-
Rear bottom for air flow of BEP Assy.
-
Left and Right side for airflow of the AC power Assy.
Figure 5-30 Air Flow Inside the Scanner
A
B
C
D