GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 4 Functional Checks
4 - 25
Section 4-8
Software Configuration Checks
Refer to Chapter 3, Section 3-6 - System Configuration for setting procedures.
Section 4-9
Peripheral Checks
Check that peripherals work as described below:
Table 4-23 Software Configuration Checks
Step
Task to do
Expected Result(s)
1.
Check Date and Time setting
Date and Time are correct
2.
Check that Location (Hospital Name and
Department) is correct
Location Name is correct
3.
Check Language setting
Language is proper
4.
Check Units setting
Units are proper
Table 4-24 Peripheral checks
Step
Task to do
Expected Result(s)
1.
Press Freeze
Stop image acquisition.
2.
Press (P1) or (P4) on the Control panel
The image displayed on the screen is printed
on B&W or Color printer depending on the
key assignment configuration.
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...